Pfitzner J
Anaesthesia. 1976 Apr;31(3):410-5. doi: 10.1111/j.1365-2044.1976.tb12336.x.
A low dead space, low resistance, non-rebreathing valve is described. When used with a constant pressure source it enables the expired gases of patients being managed on CPAP to be collected for physiological analysis. It also functions as a collect valve during normal spontaneous respiration and during IPPV.
描述了一种低死腔、低阻力、非再呼吸阀。当与恒压源一起使用时,它能收集接受持续气道正压通气(CPAP)治疗患者呼出的气体,用于生理分析。在正常自主呼吸和间歇正压通气(IPPV)期间,它还起到收集阀的作用。