Rieger A, Ebel H, Laun A
Neurochirurgische Klinik Halle, Martin-Luther-Universität.
Zentralbl Neurochir. 1995;56(1):43-8.
In 54 neurosurgical patients--suffering from intracranial tumors, hydrocephalus or different kinds of brain injury--a catheter was implanted in the lateral ventricle for continuously monitoring of the intracranial pressure (ICP). The intraventricular injection of defined amounts of sodium clearance (1-2 cm3) was followed by increased values of intracranial mean pressure--(pm) and pulse-amplitude- (PA). The recorded data showed a high significant correlation after regression analysis. After determination of the pressure-volume-index (PVI) and the theoretical pressure-volume-diagram it was examined whether there is a linear correlation between the parameters of the craniospinal space (id) reserves (constant of gradient, (PVI, K of PVD)) and the gradients (CA) of the linear PA-PM-function. A significant linear relationship, however of weak degree, could be found between these two parameters. The linear PA-PM-function, therefore, can only be understood in restricted circumstances as a direct parameter of the intracranial spacial reserves. The underlying facts of these problems are dealt within this paper.
在54例神经外科患者中——这些患者患有颅内肿瘤、脑积水或不同类型的脑损伤——在侧脑室植入一根导管,用于持续监测颅内压(ICP)。向脑室内注入一定量(1 - 2立方厘米)的钠后,颅内平均压力(pm)和脉搏振幅(PA)值升高。回归分析后,记录的数据显示出高度显著的相关性。在确定压力-容积指数(PVI)和理论压力-容积图后,研究了颅脊髓腔(id)储备参数(梯度常数,(PVI,PVD的K))与线性PA - PM函数的梯度(CA)之间是否存在线性相关性。然而,在这两个参数之间发现了显著的线性关系,但程度较弱。因此,线性PA - PM函数只能在有限的情况下被理解为颅内空间储备的直接参数。本文探讨了这些问题的潜在事实。