Englert D, Roessler N, Jeavons A, Fairless S
Packard Instrument Co., Meriden, CT 06450, USA.
Cell Mol Biol (Noisy-le-grand). 1995 Feb;41(1):57-64.
The microchannel array detector provides direct electronic detection and real time imaging of radioactive tracers in samples that traditionally have been imaged by film radioautography. The technology of the detector is described, and the main benefits of this new technology, speed and precise quantitation, are illustrated with practical results. The instrument response to a 14C dilution series (range of 10(5)) was linear over the whole range. Results from the microchannel array detector are compared to results obtained with liquid scintillation counting (LSC) and densitometry of film radioautographs. Because the instrument response is linear and consistent from day to day, it is not necessary to routinely include calibration standards as is required for film. The 14C counting efficiency was found to be essentially constant for different tissues in a whole body microtome section. As with LSC, results reported as counts per min. (CPM) are independent of counting time. Quantitative results were obtained with assays for chloramphenicol acetyl transferase in two min. and little as 0.2 DPM/mm2 of 32P was detected in seven min.
微通道阵列探测器可对样品中的放射性示踪剂进行直接电子检测和实时成像,而传统上这些样品是通过胶片放射自显影成像的。本文描述了该探测器的技术,并通过实际结果展示了这项新技术的主要优点,即速度快和定量精确。该仪器对14C稀释系列(范围为10(5))的响应在整个范围内呈线性。将微通道阵列探测器的结果与液体闪烁计数(LSC)和胶片放射自显影片密度测定的结果进行了比较。由于仪器的响应是线性的且每天一致,因此无需像胶片那样常规包含校准标准品。发现在全身切片机切片中,不同组织的14C计数效率基本恒定。与LSC一样,以每分钟计数(CPM)报告的结果与计数时间无关。在两分钟内通过氯霉素乙酰转移酶测定获得了定量结果,在七分钟内检测到低至0.2 DPM/mm2的32P。