Rithalia S V, Edwards D
Orthopaedic Mechanics Research Institute, University of Salford, UK.
J Med Eng Technol. 1994 Sep-Oct;18(5):179-81. doi: 10.3109/03091909409030249.
Non-invasive oscillometric blood pressure and pulse measured by an Omron HEM-703CP monitor were compared with arterial values obtained from direct measurements of the radial artery. An excellent correlation and agreement was found between the two methods (systolic r = 0.99; diastolic r = 0.97; pulse r = 0.99), although there was some variability among individual subjects. The range of difference between them was 0 to 10 mmHg for systolic and -6 to +5 mmHg for diastolic pressures. When tested on the bench using the Metron QA-1280 non-invasive blood pressure analyser the HEM-703CP monitor rarely exhibited errors exceeding 2-3 mmHg over a measurement range of 50-200 mmHg.