电子自卫装置对心脏起搏器的影响。

Effects of electronic autodefense devices on cardiac pacemakers.

作者信息

Moraes J C

机构信息

Biomedical Engineering Laboratory, Polytechnical Faculty, University of São Paulo, Brazil.

出版信息

Artif Organs. 1995 Mar;19(3):238-40. doi: 10.1111/j.1525-1594.1995.tb02320.x.

Abstract

The problem of interference on pacemakers as a result of extraneous electric current and electromagnetic fields is important since more than 1.5 million people worldwide have pacemakers implanted. Each year at least 100,000 people have cardiac pacemakers implanted. This article reports several effects of two electronic autodefense devices on some cardiac pacemakers. The results show that the interference affects severely the behavior of the cardiac pacemakers, but if the duration is no more than 5 s, the effects are not permanent, and the pacemakers return to the correct action as soon as the interference ceases.

摘要

由于全球有超过150万人植入了心脏起搏器,因此外部电流和电磁场对起搏器产生干扰的问题很重要。每年至少有10万人植入心脏起搏器。本文报告了两种电子自动防御装置对某些心脏起搏器的几种影响。结果表明,干扰会严重影响心脏起搏器的运行,但如果持续时间不超过5秒,影响不是永久性的,干扰一停止,起搏器就会恢复正确运行。