Phys Med Biol. 1976 May;21(3):390-8. doi: 10.1088/0031-9155/21/3/004.
As a polychromatic X-ray beam passes through matter, low energy photons are preferentially absorbed, and the (logarithmic) attenuation is no longer a linear function of absorber thickness. This leads to various artifacts in reconstructive tomography. If a water bag is used, the nonlinear attenuation in bone causes a distortion of the bone values and a spill-over inside the skull, or 'pseudo-cortex' artifact. If no water bag is used, there is an additional effect due to the varying thickness of soft tissue which causes a depression of interior values, or 'cupping'. Both artifacts can be remedied by additional prefiltering of the beam and by applying a linearization correction to the detector outputs.
当多色X射线束穿过物质时,低能光子优先被吸收,并且(对数)衰减不再是吸收体厚度的线性函数。这会在重建断层扫描中导致各种伪影。如果使用水袋,骨骼中的非线性衰减会导致骨值失真以及颅骨内出现溢出,即“假皮质”伪影。如果不使用水袋,由于软组织厚度变化会产生额外影响,导致内部值降低,即“杯状”伪影。通过对射线束进行额外的预滤波以及对探测器输出应用线性化校正,可以校正这两种伪影。