Suppr超能文献

Application of an electrochemical detector with a graphite electrode to liquid chromatographic determination of penicillamine and captopril in biological samples.

作者信息

Wakabayashi H, Yamato S, Nakajima M, Shimada K

机构信息

Department of Analytical Chemistry, Niigata College of Pharmacy, Japan.

出版信息

J Pharm Biomed Anal. 1994 Sep;12(9):1147-52. doi: 10.1016/0731-7085(94)00072-7.

Abstract

A high-performance liquid chromatographic determination of penicillamine and captopril in rat serum, liver and kidney samples is described. An electrochemical detector with a graphite working electrode at a potential of +0.9 V vs the Ag/AgCl reference electrode is used for the detection system. Linear responses of the peak height to the amount of samples injected were obtained in a range of 0.1-500 ng on-column and 0.5-500 ng on-column for penicillamine and captopril with correlation coefficients of 0.997 and 0.995, respectively. Detection limits at a signal-to-noise ratio of 3 were 20 and 300 pg for penicillamine and captopril, respectively. The graphite electrode has a long lifetime of about 4 months with continuous use, even with the high voltage supplied. The analytical application of this method to the determination of penicillamine and captopril in biological samples was successful.