The glutamate biosensor and its application to flow injection analysis system.

作者信息

Ye B, Li Q, Li Y, Yu J

机构信息

Research Institute of Biochemical Engineering, East China University of Science and Technology, Shanghai.

出版信息

Chin J Biotechnol. 1994;10(2):83-9.

PMID:7803693
Abstract

A micro-enzyme electrode was fabricated by cross-linking L-glutamate oxidase with glutaraldehyde on aminopropyl-platinized platinum wire. A flow injection analysis system with glutamate sensor was used for L-glutamate determination. The peak current is linearly related to the L-glutamate concentration in the range of 0.02-2.0 mM, with good performance, accuracy (CV = 0.4%), fast response (< 60s), and stability (> 20 days). The system was applied to determine the concentration of L-glutamate in a fermentation broth. The recovery rate was in the range of 98.7-107.5%.

摘要

通过在氨丙基镀铂铂丝上用戊二醛交联L-谷氨酸氧化酶制备了一种微酶电极。使用带有谷氨酸传感器的流动注射分析系统测定L-谷氨酸。在0.02 - 2.0 mM范围内,峰电流与L-谷氨酸浓度呈线性关系,具有良好的性能、准确性(CV = 0.4%)、快速响应(< 60秒)和稳定性(> 20天)。该系统用于测定发酵液中L-谷氨酸的浓度。回收率在98.7 - 107.5%范围内。