Fried D W, DeBenedetto B N, Zombolas T L, Leo J J
Albert Einstein Medical Center, Philadelphia, Pennsylvania.
Perfusion. 1994;9(5):363-72. doi: 10.1177/026765919400900509.
The purpose of this study was to clinically evaluate the degree of improvement, if any, in the oxygen transfer performance of the recently released Medtronic Maxima Plus membrane oxygenator. The outside diameter of the hollow fibres was reduced, increasing the membrane surface area from 2.0 m2 to 2.3 m2 without altering the polycarbonate housing. Maximum extrapolated oxygen transfer of the Maxima Plus (444 ml O2/minute) was increased 23.68% when compared to the Maxima (359 ml O2/minute). When expressed per metre squared of membrane surface area, the Maxima Plus had an increase of 13.5 ml O2/m2/minute (7.24%) over the Maxima. Pressure drop across the Maxima Plus was within 3.5 mmHg of the Maxima over the range of clinical blood flows indicating that the fibre bundle packing density was not significantly altered. Oxygen transfer consistency, expressed as a function of the standard deviation of oxygenator performance index values, was not significantly different for the two oxygenators. We concluded that the improvement in total oxygen transfer was due to an increase in membrane surface area as well as enhanced transfer efficiency per metre squared. We believe that the improved oxygen transfer performance was accomplished without impacting significantly upon the other attributes of the oxygenator (e.g., pressure drop, consistency, priming volume).
本研究的目的是临床评估最新推出的美敦力Maxima Plus膜式氧合器的氧传递性能是否有所改善(若有改善,则评估改善程度)。中空纤维的外径减小,在不改变聚碳酸酯外壳的情况下,膜表面积从2.0平方米增加到2.3平方米。与Maxima(359毫升氧气/分钟)相比,Maxima Plus的最大外推氧传递量(444毫升氧气/分钟)增加了23.68%。以每平方米膜表面积计算,Maxima Plus比Maxima增加了13.5毫升氧气/平方米/分钟(7.24%)。在临床血流范围内,Maxima Plus的压降在Maxima的3.5毫米汞柱以内,这表明纤维束的填充密度没有显著改变。以氧合器性能指标值的标准差表示的氧传递一致性,两种氧合器之间没有显著差异。我们得出结论,总氧传递的改善是由于膜表面积增加以及每平方米传递效率提高。我们认为,在不显著影响氧合器其他特性(如压降、一致性、预充量)的情况下,实现了氧传递性能的改善。