Walters D L, James J E, Vest F B, Karnes H T
Battelle, Columbus, OH 43201.
Biomed Chromatogr. 1994 Sep-Oct;8(5):207-11. doi: 10.1002/bmc.1130080503.
Fluorescence and chemiluminescence detection were compared for HPLC analysis of the fluorescamine derivative of histamine. The kinetic behaviour of the chemiluminescent response for the derivative was characterized in a static system. An HPLC method was optimized for the derivative using fluorescence detection. Fluorescence detection was linear over the range of 166-1666 pg on column for the fluorescamine-histamine derivative with a limit of detection of 13 pg on column. Using a detector designed for optimal use with chemiluminescence, the chemiluminescence response of the fluorescamine derivative was linear over a range of 1.66-16.6 ng on column with a limit of detection of 1.0 ng on column. These results exemplify a case in which superior detectibility is provided by fluorescence over chemiluminescence, and contradicts many reports comparing fluorescence to chemiluminescence. The authors conclude that chemiluminescence should be considered when indicated by conditions established for separation that are favourable for the observation of chemiluminescence. These conditions include sufficiently low excitation energies corresponding to an excitation maximum greater than 400 nm, favourable dipole character of analytes, mobile phases of high organic content, and an appropriate pH of the mobile phase.
对组胺的荧光胺衍生物进行高效液相色谱(HPLC)分析时,比较了荧光检测和化学发光检测。在静态系统中对该衍生物的化学发光响应动力学行为进行了表征。使用荧光检测对该衍生物的HPLC方法进行了优化。对于荧光胺 - 组胺衍生物,柱上荧光检测在166 - 1666 pg范围内呈线性,柱上检测限为13 pg。使用专为与化学发光最佳配合设计的检测器,荧光胺衍生物的化学发光响应在柱上1.66 - 16.6 ng范围内呈线性,柱上检测限为1.0 ng。这些结果例证了一种情况,即荧光检测在可检测性方面优于化学发光检测,这与许多比较荧光和化学发光的报告相矛盾。作者得出结论,当分离条件有利于化学发光观察时,应考虑使用化学发光检测。这些条件包括对应于大于400 nm的激发最大值的足够低的激发能量、分析物有利的偶极特性、高有机含量的流动相以及流动相的适当pH值。