Tengborn L, Palmblad S, Wojciechowski J, Peterson L E, Stigendal L
Department of Medicine, Sahlgren's Hospital, Göteborg, Sweden.
Haemostasis. 1994 Nov-Dec;24(6):344-50. doi: 10.1159/000217125.
In order to evaluate the usefulness of determining D-dimer and thrombin-antithrombin III complex (TAT) in the diagnosis of deep venous thrombosis (DVT), three D-dimer assays were tried: one ELISA and one latex test from Diagnostica Stago and one new latex method from Biopool. TAT was assessed using an ELISA (Behringwerke). We studied 96 consecutive outpatients with suspected DVT, of whom 36 had phlebographically confirmed DVT. Statistical calculations showed high sensitivity and a negative predictive value for the D-dimer ELISA (97% for both), confirming the results obtained by others. The new latex method (Biopool) showed similar figures (96% for both). The latex method from Diagnostica Stago and TAT showed lower sensitivity and negative predictive values. No differences in the D-dimer results were found with or without antifibrinolytics in the tubes for blood sampling. Our data suggest that negative results when using the new simple and cheap latex method (Biopool) may reduce the number of phlebographic examinations.
为了评估测定D - 二聚体和凝血酶 - 抗凝血酶III复合物(TAT)在诊断深静脉血栓形成(DVT)中的实用性,尝试了三种D - 二聚体检测方法:一种是Diagnostica Stago公司的酶联免疫吸附测定(ELISA)和乳胶试验,另一种是Biopool公司的新型乳胶法。TAT采用ELISA法(拜耳公司产品)进行评估。我们研究了96例连续的疑似DVT门诊患者,其中36例经静脉造影证实患有DVT。统计计算表明,D - 二聚体ELISA具有高敏感性和阴性预测值(两者均为97%),证实了其他人获得的结果。新型乳胶法(Biopool)显示出相似的数据(两者均为96%)。Diagnostica Stago公司的乳胶法和TAT显示出较低的敏感性和阴性预测值。在用于采血的试管中,无论是否使用抗纤维蛋白溶解剂,D - 二聚体结果均未发现差异。我们的数据表明,使用新型简单且廉价的乳胶法(Biopool)得到的阴性结果可能会减少静脉造影检查的次数。