Krauss J K, Droste D W
Department of Neurosurgery, University of Freiburg, Germany.
Neurol Res. 1994 Oct;16(5):398-402. doi: 10.1080/01616412.1994.11740261.
Intracranial pressure (ICP) was monitored continuously for one night in 36 patients with suspected symptomatic normal pressure hydrocephalus (NPH) to identify patients who might benefit from subsequent shunting. In 33 of these patients middle cerebral artery (MCA) blood flow velocity by means of transcranial Doppler sonography (TCD) and ICP were recorded simultaneously. ICP B-waves always paralleled changes in the TCD signal (TCD B-wave equivalents). The relative frequency of ICP B-waves was predictable by TCD, albeit slightly underestimated due to a generally lower relative amplitude of the TCD B-wave equivalents. However, the same TCD B-wave equivalent amplitude could be accompanied by quite different ICP changes in different patients. Considering the baseline values in the absence of pressure waves, there was no significant relationship between ICP and TCD resistance index (Pourcelot) in different patients. Raising ICP by injection of 10 ml saline into the ventricle, however was accompanied by an increased TCD resistance index in the individual patient. As the relative frequency of B-wave activity is assumed to be an indicator for shunt responsiveness, continuous TCD monitoring can be used as a screening procedure to detect the presence and the relative frequency of B-wave activity in patients with suspected NPH. However, since neither the absolute ICP nor the amplitude of spontaneous oscillations can be predicted, TCD monitoring is not suitable to replace ICP monitoring.
对36例疑似有症状性正常压力脑积水(NPH)的患者连续监测了一晚的颅内压(ICP),以确定可能从后续分流术中获益的患者。在其中33例患者中,通过经颅多普勒超声(TCD)同时记录大脑中动脉(MCA)血流速度和ICP。ICP的B波总是与TCD信号的变化(TCD B波等效物)平行。ICP B波的相对频率可由TCD预测,尽管由于TCD B波等效物的相对幅度通常较低而略有低估。然而,相同的TCD B波等效物幅度在不同患者中可能伴随着相当不同的ICP变化。考虑无压力波时的基线值,不同患者的ICP与TCD阻力指数(普尔塞洛指数)之间无显著相关性。然而,向脑室内注射10 ml生理盐水使ICP升高时,个体患者的TCD阻力指数会增加。由于B波活动的相对频率被认为是分流反应性的指标,连续TCD监测可作为一种筛查程序,用于检测疑似NPH患者中B波活动的存在及其相对频率。然而,由于既无法预测绝对ICP,也无法预测自发振荡的幅度,TCD监测不适于替代ICP监测。