Clinically feasible stable isotope technique at a reasonable price: analysis of 13CO2/12CO2-abundance in breath samples with a new isotope selective-nondispersive infrared spectrometer.

作者信息

Braden B, Haisch M, Duan L P, Lembcke B, Caspary W F, Hering P

机构信息

Medical Department II, Center of Internal Medicine, Johann Wolfgang Goethe-University Frankfurt/Main.

出版信息

Z Gastroenterol. 1994 Dec;32(12):675-8.

PMID:7871857
Abstract

Up to now, stable isotope analysis of carbon dioxide in breath samples is carried out with sensitive but very expensive and complex isotope ratio mass spectrometry (IRMS). Aiming at a more widespread application of breath tests in gastroenterological diagnostic routine, we tested a newly developed isotope selective non-dispersive infrared spectrometer (NDIRS) in comparison to IRMS. 13C-urea breath tests were performed in 63 patients as the routine screening method for Helicobacter pylori infection. Breath samples at baseline and (15) 30 min after administration of the test solution containing 13C-urea were analysed both by NDIRS and conventional IRMS. The correlation between the delta values of both devices was linear and in good agreement (r = 0.96; p < 0.0001; Y = 1.01 X -0.94). Comparing the delta over baseline-values, the correlation was Y = 1.11 X -0.36 (r = 0.98; p < 0.0001). Referring to the diagnosis of Helicobacter pylori infection with IRMS we calculated a sensitivity of 95.0% and an unchanged specificity (100%) for NDIR analysis. In conclusion, NDIRS appears a promising, easy to operate, and low cost potential alternative to conventional IRMS thus encouraging further detailed investigation and more widespread application of the noninvasive stable isotope technique in breath tests for gastrointestinal function testing.

摘要

到目前为止,呼气样本中二氧化碳的稳定同位素分析是使用灵敏但非常昂贵且复杂的同位素比率质谱仪(IRMS)进行的。为了使呼气测试在胃肠病诊断常规中得到更广泛的应用,我们将一种新开发的同位素选择性非色散红外光谱仪(NDIRS)与IRMS进行了比较测试。对63例患者进行了¹³C尿素呼气试验,作为幽门螺杆菌感染的常规筛查方法。通过NDIRS和传统的IRMS对基线时以及服用含¹³C尿素的测试溶液后30分钟的呼气样本进行了分析。两种设备的δ值之间的相关性呈线性且一致性良好(r = 0.96;p < 0.0001;Y = 1.01X - 0.94)。比较相对于基线值的δ值,相关性为Y = 1.11X - 0.36(r = 0.98;p < 0.0001)。参照IRMS对幽门螺杆菌感染的诊断,我们计算出NDIR分析的灵敏度为95.0%,特异性不变(100%)。总之,NDIRS似乎是一种有前景、易于操作且低成本的传统IRMS替代方法,因此鼓励对这种非侵入性稳定同位素技术在胃肠功能测试呼气试验中的进一步详细研究和更广泛应用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验