Suppr超能文献

Quantitative water mapping of cryosectioned cells by electron energy-loss spectroscopy.

作者信息

Sun S Q, Shi S L, Hunt J A, Leapman R D

机构信息

Biomedical Engineering and Instrumentation Program, National Center for Research Resources, NIH, Bethesda, MD 20892.

出版信息

J Microsc. 1995 Jan;177(Pt 1):18-30. doi: 10.1111/j.1365-2818.1995.tb03530.x.

Abstract

A direct technique based on electron energy-loss spectroscopy (EELS) in the scanning transmission electron microscope (STEM) has been developed to map subcellular distributions of water in frozen-hydrated biological cryosections. Previously, methods for water determination have been indirect in that they have required the cryosections to be dehydrated first. The new approach makes use of spectrum-imaging, where EELS data are collected in parallel at each pixel. Several operations are required to process the spectra including: subtraction of the detector dark current, deconvolution by the detector point-spread function, removal of plural inelastic scattering and correction for the support film. The resulting single scattering distributions are fitted to standard reference spectra at each pixel, and water content can be determined from the fitting coefficients. Although the darkfield or brightfield image from a hydrated cryosection shows minimal structure, the processed EELS image reveals strong contrast due to variations in water content. Reference spectra have been recorded from the major biomolecules (protein, lipid, carbohydrate, nucleic acid) as well as from vitrified water and crystalline ice. It has been found that quantitative results can be obtained for the majority of subcellular compartments by fitting only water and protein reference spectra, and the accuracy of the method for these compartments has been estimated as +/- 3.5%. With the present instrumentation the maximum allowed dose of 2 x 10(3) e/nm2 limits the useful spatial resolution to around 80 nm for +/- 5% precision at a single pixel. By averaging pixel intensities a value of 56.8% with a precision of +/- 2.0% has been determined for the water content of liver mitochondria. The water mapping technique may prove useful for applications to cell physiology and pathophysiology.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验