10 kHz microsecond pulsed X-ray generator utilising a hot-cathode triode with variable durations for biomedical radiography.

作者信息

Sato E, Sagae M, Takahashi K, Shikoda A, Oizumi T, Hayasi Y, Tamakawa Y, Yanagisawa T

机构信息

Department of Physics, Iwate Medical University, Morioka, Japan.

出版信息

Med Biol Eng Comput. 1994 May;32(3):295-301. doi: 10.1007/BF02512525.

Abstract

A 10 kHz pulsed X-ray generator utilising a hot-cathode triode in conjunction with a new type of grid control device for controlling X-ray duration is described. The energy-storage condenser was charged up to 70 kV by a power supply, and the electric charges in the condenser were discharged to the X-ray tube repetitively by the grid control device. The maximum values of the grid voltage (negative value), the tube voltage, and the tube current were -1.5 kV, 70 kV, and 0.4 A, respectively. The duration of the flash X-ray pulse was primarily determined by the time constant of the grid control device and the cut-off voltage of thermoelectrons. The X-ray duration was controlled within a region of less than 1 ms; the X-ray intensity with a pulse width of 0.27 ms, a charged voltage of 70 kV, and a peak tube current of 0.4 A was 0.92 microC kg-1 at 0.5 m per pulse. The maximum repetition rate was about 10 kHz, and the size of the focal spot was about 3.5 x 3.5 mm.

摘要

描述了一种10kHz脉冲X射线发生器,它利用热阴极三极管与一种新型栅极控制装置相结合来控制X射线持续时间。储能电容器由电源充电至70kV,电容器中的电荷由栅极控制装置重复放电至X射线管。栅极电压(负值)、管电压和管电流的最大值分别为-1.5kV、70kV和0.4A。闪光X射线脉冲的持续时间主要由栅极控制装置的时间常数和热电子的截止电压决定。X射线持续时间控制在小于1ms的范围内;在每个脉冲0.5m处,脉冲宽度为0.27ms、充电电压为70kV、峰值管电流为0.4A时的X射线强度为0.92微库仑每千克。最大重复率约为10kHz,焦点尺寸约为3.5×3.5mm。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验