He Q, Viljanen M K, Mertsola J
National Public Health Institute, Department in Turku, Finland.
Mol Cell Probes. 1994 Apr;8(2):155-9. doi: 10.1006/mcpr.1994.1021.
The effects of thermocyclers and primers on the reproducibility of banding patterns in randomly amplified polymorphic DNA analysis were tested. Purified Bordetella pertussis DNA was analysed with four primers (12-mer), which did not differ from each other in their GC-content. Three different thermocycler models from two manufacturers were tested. Three of the primers produced consistent banding patterns in separate wells of the reaction plates of individual thermocyclers, and in different thermocycler models. In contrast, the fourth primer showed extensive between-instrument variation, and the within-instrument variation with this primer was acceptable only when the most advanced thermocycler model was used. We conclude that only primers that provide consistent banding patterns in different environments should be chosen for interlaboratory use in randomly amplified polymorphic DNA analysis. The primers showing variability in banding patterns can only be used if they produce consistent results in the thermocycler used.
测试了热循环仪和引物对随机扩增多态性DNA分析中条带模式重现性的影响。使用四种引物(12聚体)对纯化的百日咳博德特氏菌DNA进行分析,这些引物的GC含量彼此无差异。测试了来自两个制造商的三种不同型号的热循环仪。其中三种引物在各个热循环仪反应板的不同孔中以及在不同型号的热循环仪中产生了一致的条带模式。相比之下,第四种引物显示出仪器间的广泛差异,并且只有在使用最先进的热循环仪型号时,该引物在仪器内的差异才是可接受的。我们得出结论,在随机扩增多态性DNA分析中,只有在不同环境中能提供一致条带模式的引物才应选择用于实验室间使用。条带模式显示出变异性的引物只有在其使用的热循环仪中产生一致结果时才能使用。