The effect of position on magnetic resonance evaluation of cerebrospinal fluid shunt function.

作者信息

Frank E, Buonocore M, Hein L

机构信息

Division Neurosurgery, Oregon Health Sciences University, Portland 97201.

出版信息

Neurol Res. 1994 Jun;16(3):168-70. doi: 10.1080/01616412.1994.11740220.

Abstract

Since cerebrospinal fluid shunt malfunction is common and its diagnosis often requires invasive testing, we have evaluated magnetic resonance imaging (MRI) as a noninvasive test for shunt function. In prior experiments, MRI has been shown to be capable of detecting flows as low as 0.4 cc h-1 in shunt systems. Because patient movement and position in an MRI scanner are variable, we have studied the effects of shunt position on the ability to reliably detect the rate of CSF flow. Flow related enhancement of fluid flowing from 0 to 2 cc h-1 was measured in a specialized section of tubing using multiple spin echo scans (TR2000,TE20). The angle of the specialized tubing was varied relative to the axis of the MRI scanner and then the intensity of the MR signal at specific points in the tubing was measured at different flow rates. No flow standards were used as controls. The results show that the linear relationship between signal intensity and flow that had previously been demonstrated was significantly altered by changes of position. Positions angled superior to the axis of the scanner enhanced the signal while lateral changes in the plane of the scan degraded the signal. These results suggest that in order for MRI to become an accurate test for shunt function, techniques must be developed to overcome the effects of shunt position.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验