Comparison of transit-time and Doppler ultrasound methods for measurement of flow in aortocoronary bypass grafts during cardiac surgery.

作者信息

Matre K, Birkeland S, Hessevik I, Segadal L

机构信息

Medical Department A, University of Bergen, Haukeland Hospital, Norway.

出版信息

Thorac Cardiovasc Surg. 1994 Jun;42(3):170-4. doi: 10.1055/s-2007-1016481.

Abstract

To evaluate the accuracy of flow measurements in aortocoronary bypass grafts with the ultrasound transit-time method, an in vitro and in vivo comparison was carried out. The in vitro comparison with evaluation against both true flow and the ultrasound Doppler method, was carried out with a fresh saphenous vein mounted in a pulsatile flow rig. The two flow probes were placed on the graft 4-5 cm apart to avoid acoustic interference, and blood was pumped through the system at different flow rates. The comparison between the methods showed excellent agreement with a linear correlation coefficient of 0.996, and a mean error of -2.9 ml/min with limits of agreement +/- 13.1 ml/min (+/- 2 SD = 95% of measured differences between these limits). However, against true flow, both methods overestimated flow slightly with mean error 4.4 and 7.3 ml/min for the transit-time and Doppler, respectively. Both methods showed excellent correlation with true flow (correlation coefficient 0.998 for the transit-time and 0.997 for the Doppler method). The in vivo accuracy was evaluated by comparison of the two methods in 9 patients. The two probes were placed on the same saphenous vein grafts 4-5 cm apart, and a total of 34 measurements in 17 grafts were carried out including measurements at baseline and after papaverine injection. The correlation coefficient was 0.990 and linear regression analysis gave the equation: Transit-time flow = 1.00 x Doppler flow + 1.3. In terms of flow, the mean error was 1.5 ml with limits of agreement +/- 17.2 ml.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

为了用超声渡越时间法评估主动脉冠状动脉搭桥术中血流测量的准确性,进行了一项体外和体内比较。体外比较是在一个搏动流装置中安装新鲜大隐静脉,针对真实血流和超声多普勒法进行评估。将两个血流探头放置在移植物上,相距4 - 5厘米以避免声学干扰,并以不同流速将血液泵入系统。方法之间的比较显示出极好的一致性,线性相关系数为0.996,平均误差为-2.9毫升/分钟,一致性界限为±13.1毫升/分钟(±2标准差 = 这些界限之间测量差异的95%)。然而,与真实血流相比,两种方法都略微高估了血流,渡越时间法和多普勒法的平均误差分别为4.4和7.3毫升/分钟。两种方法与真实血流都显示出极好的相关性(渡越时间法的相关系数为0.998,多普勒法为0.997)。通过在9名患者中比较两种方法来评估体内准确性。将两个探头放置在同一条大隐静脉移植物上,相距4 - 5厘米,在17个移植物中总共进行了34次测量,包括基线测量和罂粟碱注射后的测量。相关系数为0.990,线性回归分析得出方程:渡越时间血流 = 1.00×多普勒血流 + 1.3。就血流而言,平均误差为1.5毫升,一致性界限为±17.2毫升。(摘要截短于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验