Automated evaluation of hip axis length for predicting hip fracture.

作者信息

Faulkner K G, McClung M, Cummings S R

机构信息

Providence Center for Osteoporosis Research, Portland, Oregon.

出版信息

J Bone Miner Res. 1994 Jul;9(7):1065-70. doi: 10.1002/jbmr.5650090714.

Abstract

The hip axis length has been shown in previous studies to be predictive of hip fracture independent of age and femoral bone density. The first studies of hip axis length were performed by manual measurement of dual x-ray absorptiometry (DXA) scan printouts. In this study, an automated analysis procedure is defined using software tools provided by the DXA manufacturer. Manual and automatic hip axis length measurements in 198 women were highly correlated (r = 0.98). Because of scaling factors of the printout, the automatic measurement was 58% longer than the manual value. Precision of the automatic measurement, based upon triplicate DXA scans of 33 women, was 0.07 cm or 0.68%. To define normative data, the hip axis length was measured from femoral DXA scans of 471 female volunteers aged 40-92 scanned on 14 different Hologic QDR-1000 systems. Mean hip axis length was 10.5 cm, with a standard deviation of 0.62 cm. No significant relationship between hip axis length and age was found (r = 0.07, P = 0.15). Based on previously reported odds ratios corrected for femoral bone density, age, height, and weight, an automatic hip axis length measurement of 11.0 cm is associated with a twofold increase in hip fracture risk compared with a woman with an average hip axis length. A hip axis length value of 11.6 cm increases hip fracture risk by a factor of 4 compared with a woman with a normal hip dimension. We conclude that the hip axis length can be easily incorporated into existing DXA hip analysis software in combination with a bone density measurement.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

以往研究表明,髋轴长度可独立于年龄和股骨骨密度预测髋部骨折。最早关于髋轴长度的研究是通过手动测量双能X线吸收法(DXA)扫描打印件进行的。在本研究中,利用DXA设备制造商提供的软件工具定义了一种自动分析程序。对198名女性进行的手动和自动髋轴长度测量具有高度相关性(r = 0.98)。由于打印件的缩放因素,自动测量值比手动测量值长58%。基于对33名女性进行的三次重复DXA扫描,自动测量的精度为0.07 cm或0.68%。为确定正常数据,对在14台不同的Hologic QDR - 1000系统上扫描的471名年龄在40 - 92岁的女性志愿者的股骨DXA扫描图像测量髋轴长度。平均髋轴长度为10.5 cm,标准差为0.62 cm。未发现髋轴长度与年龄之间存在显著关系(r = 0.07,P = 0.15)。根据先前报道的经股骨骨密度、年龄、身高和体重校正的优势比,与髋轴长度平均的女性相比,自动测量的髋轴长度为11.0 cm时,髋部骨折风险增加两倍。与髋部尺寸正常的女性相比,髋轴长度值为11.6 cm时,髋部骨折风险增加4倍。我们得出结论,髋轴长度可轻松整合到现有的DXA髋部分析软件中,并结合骨密度测量。(摘要截短为250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验