Cazalaà J B, Fougère S, Barrier G
Département d'Anesthésie-Réanimation, Hôpital Necker-Enfants Malades, Paris.
Ann Fr Anesth Reanim. 1994;13(3):336-49. doi: 10.1016/s0750-7658(94)80042-1.
This study compared 45 electric infusion devices, either constant flow pumps or syringe pumps, available in France. These devices were tested for start-up delay, time to alarm following occlusion, bolus following occlusion, accuracy of flow rate and pressure output. Ergonomic, environmental and various technical criteria such as safety in continuous perfusion, correct positioning of infusion sets, air in line and end of perfusion detection were also considered. Users can compare these devices in easy to read tables. The newest devices include systems improving the quality of flow.
本研究对比了法国市面上的45种电动输注设备,包括恒流泵或注射泵。对这些设备进行了启动延迟、阻塞后报警时间、阻塞后推注、流速准确性和压力输出等方面的测试。还考虑了人体工程学、环境以及各种技术标准,如连续灌注安全性、输液装置正确定位、管路中的空气和灌注结束检测。用户可以通过易于阅读的表格比较这些设备。最新的设备包括提高流速质量的系统。