Scatter-reduction characteristics of an infinity-focused gridded radiographic cassette.

作者信息

Floyd C E, Chotas H G, Ravin C E

机构信息

Department of Radiology, Duke University, Durham, North Carolina 27710.

出版信息

Invest Radiol. 1994 Sep;29(9):852-5. doi: 10.1097/00004424-199409000-00010.

Abstract

RATIONALE AND OBJECTIVES

The scatter-reduction properties of a new infinity-focused grid incorporated into a radiographic imaging cassette were analyzed.

METHODS

A polystyrene chest phantom was imaged using the cassette and bedside radiographic procedure. Scatter fractions were measured using a beam-stop technique at several anatomically equivalent locations. The performance of this cassette also was evaluated as a function of the orientation angle.

RESULTS

The gridded cassette provided a decrease in scatter fraction from 61% to 49% in the lung and from 87% to 76% in the mediastinum. To obtain equivalent film density when the grid was used, the exposure was increased by a factor of 3.0. While there was a decrease in scatter clean-up as the film cassette was tilted from the perpendicular, the grid performed well out to an angle of 10 degrees.

CONCLUSION

The gridded cassette for portable radiography provides scatter reduction with little sensitivity to alignment. The availability of this device could improve scatter rejection, and therefore, contrast in portable radiographic imaging.

摘要

相似文献

1
Scatter-reduction characteristics of an infinity-focused gridded radiographic cassette.
Invest Radiol. 1994 Sep;29(9):852-5. doi: 10.1097/00004424-199409000-00010.
2
Rare-earth scatter fractions in chest radiography.
Med Phys. 1993 Mar-Apr;20(2 Pt 1):475-9. doi: 10.1118/1.597041.
3
[Examination for Effectiveness of Scatter Correction in Portable Chest Radiography].
Nihon Hoshasen Gijutsu Gakkai Zasshi. 2016;72(12):1207-1215. doi: 10.6009/jjrt.2016_JSRT_72.12.1207.
4
Laser alignment system for high-quality portable radiography.
Radiographics. 1992 Jan;12(1):111-20. doi: 10.1148/radiographics.12.1.1734456.
5
Evaluation of a parallel hole grid for bedside chest imaging.
Invest Radiol. 1994 Jul;29(7):682-8. doi: 10.1097/00004424-199407000-00004.
6
Dose-image quality optimisation in digital chest radiography.
Radiat Prot Dosimetry. 2005;114(1-3):269-72. doi: 10.1093/rpd/nch546.
7
Clinical evaluation of twin screen-film chest radiography: cost effective lung and mediastinal imaging.
Eur J Radiol. 1998 Mar;27(1):61-6. doi: 10.1016/s0720-048x(97)00031-4.
8
Effectiveness of antiscatter grids in digital radiography. A phantom study.
Invest Radiol. 1994 Jun;29(6):636-42. doi: 10.1097/00004424-199406000-00007.
10
Technical aspects of twin screen-film chest radiography: cost effective lung and mediastinal imaging.
Eur J Radiol. 1998 Mar;27(1):53-60. doi: 10.1016/s0720-048x(97)00033-8.

引用本文的文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验