Updike S J, Shults M C, Rhodes R K, Gilligan B J, Luebow J O, von Heimburg D
Department of Medicine, University of Wisconsin Center for Health Sciences, Madison, WI 53792.
ASAIO J. 1994 Apr-Jun;40(2):157-63.
We studied the long-term in vitro and in vivo performance of enzyme electrode glucose sensors. Single commercially produced enzyme-active membranes remained functional for estimating glucose in vitro for 14-36 months. These membranes were implanted subcutaneously in rats for 1 year and, upon explanation, remained functional for measuring glucose in vitro. Sensors with these membranes plus an additional outer membrane with lower glucose permeability allowed glucose monitoring in the low oxygen tension of subcutaneous tissue. These sensors were surgically implanted in three nondiabetic dogs. Each sensor implant was coupled to a radio transmitter to allow continuous long-term glucose monitoring in these awake unrestrained dogs. In vivo sensor performance was evaluated by intravenous glucose infusion, with reference blood glucose determinations made in the clinical laboratory. These subcutaneously implanted sensors tracked changes in plasma glucose for up to 12 weeks. The in vivo initial response for three sensor implants was approximately 35 sec (n = 8). Sensor peak response to glucose after bolus infusion ranged from 3 to 14 min. Stability of sensor sensitivity within +/- 15% for more than 1 month was demonstrated in two of the dogs. Sensor lifetime was limited not by loss of enzyme activity, but by biodegradation of the outermost polyurethane membrane. The findings suggest that long-term continuous monitoring of blood glucose using a subcutaneously implanted enzyme electrode sensor may be possible.
我们研究了酶电极葡萄糖传感器的长期体外和体内性能。单个商业生产的酶活性膜在体外估计葡萄糖时可保持功能14至36个月。将这些膜皮下植入大鼠体内1年,解剖后,其在体外测量葡萄糖时仍保持功能。带有这些膜以及额外一层葡萄糖渗透性较低的外膜的传感器,能够在皮下组织的低氧张力环境中进行葡萄糖监测。将这些传感器通过手术植入三只非糖尿病犬体内。每个传感器植入物都与一个无线电发射器相连,以便在这些清醒且不受束缚的犬身上进行连续长期的葡萄糖监测。通过静脉输注葡萄糖来评估体内传感器的性能,并在临床实验室测定参考血糖。这些皮下植入的传感器能够追踪血浆葡萄糖变化长达12周。三个传感器植入物的体内初始反应约为35秒(n = 8)。推注葡萄糖后传感器的峰值反应时间为3至14分钟。在其中两只犬身上证明了传感器灵敏度在超过1个月的时间内稳定在±15%以内。传感器的使用寿命不是受酶活性丧失的限制,而是受最外层聚氨酯膜生物降解的限制。这些发现表明,使用皮下植入的酶电极传感器对血糖进行长期连续监测或许是可行的。