Hart J D, Withers R T, Tucker R C
Exercise Physiology Laboratory, School of Education, Flinders University of South Australia, Bedford Park.
Eur Respir J. 1994 Apr;7(4):813-6. doi: 10.1183/09031936.94.07040813.
The P.K. Morgan Mark II Ventilometer comprises a turbine transducer and a control/readout unit, which contains either a VENTX5A or VENTX6A programme. This equipment is used to measure minute ventilation during oxygen consumption tests. The precision and accuracy of six of these devices (a VENTX5A with two turbines, and two VENTX6As each with two different turbines) were examined during continuous and pulsatile flows; accuracy was affected by the syringe rate during the calibration procedure. The reference standard was a 350 l Collins chain-compensated gasometer. At continuous flows, the two VENTX5A devices showed accuracies of 99.5-100.5% over the range 60-160 l.min-1 whereas those for the four VENTX6As were 99.3-102.1% over the range 60-200 l.min-1. The mean accuracies of sinusoidal flows for the VENTX5A ranged between 96.1-101.3% for minute volumes spanning 15-200 l.min-1; comparable data for the VENTX6As were 95.9-98.9%. These mean accuracies improved to 98.1-101.3% for the VENTX5A at measured minute volumes > 80 l.min-1. Omission of the two lowest tidal volumes (1 and 1.5 l) for the VENTX6As narrowed the range of accuracy to 96.9-98.5%. The coefficients of variation for five trials at each minute ventilation were 0-0.2%. We conclude that, while the P.K. Morgan Mark II Ventilometer has excellent precision, its accuracy is dependent on the calibration syringe rate, minute ventilation and program.
P.K.摩根Mark II通气计由一个涡轮传感器和一个控制/读数单元组成,该单元包含VENTX5A或VENTX6A程序。此设备用于在耗氧测试期间测量分钟通气量。在连续流和脉动流期间,对其中六个设备(一个配备两个涡轮的VENTX5A以及两个分别配备两个不同涡轮的VENTX6A)的精度和准确性进行了检查;在校准过程中,准确性受注射器速率的影响。参考标准是一台350升的柯林斯链式补偿气量计。在连续流情况下,两个VENTX5A设备在60 - 160升·分钟⁻¹范围内的准确率为99.5 - 100.5%,而四个VENTX6A设备在60 - 200升·分钟⁻¹范围内的准确率为99.3 - 102.1%。对于VENTX5A,在分钟通气量为15 - 200升·分钟⁻¹的正弦流中,平均准确率在96.1 - 101.3%之间;VENTX6A的可比数据为95.9 - 98.9%。对于VENTX5A,在测量的分钟通气量> 80升·分钟⁻¹时,这些平均准确率提高到98.1 - 101.3%。对于VENTX6A,省略两个最低潮气量(1升和1.5升)后,准确率范围缩小至96.9 - 98.5%。每分钟通气量进行五次试验的变异系数为0 - 0.2%。我们得出结论,虽然P.K.摩根Mark II通气计具有出色的精度,但其准确性取决于校准注射器速率、分钟通气量和程序。