Ordidge R J, Helpern J A, Qing Z X, Knight R A, Nagesh V
NIH Center for Stroke Research, Department of Neurology, Henry Ford Health Sciences Center, Detroit, MI 48202.
Magn Reson Imaging. 1994;12(3):455-60. doi: 10.1016/0730-725x(94)92539-9.
Patient motion can seriously degrade the quality of diffusion-weighted MR images obtained using standard 2DFT imaging procedures. The main source of error arises from an MR signal phase-shift error which is proportional to the magnitude of the motion. A modified pulse sequence is proposed which uses the phase information from an additional spin echo to correct for patient motion. Application of this technique is demonstrated for a human brain study, which greatly improves the quantification of diffusion values from regions of brain tissue.