Fujii T, Kurihara Y, Arimoto H, Mitsutsuka Y
National Institute for Environmental Studies, Ibaraki, Japan.
Anal Chem. 1994 Jun 1;66(11):1884-9. doi: 10.1021/ac00083a018.
Surface ionization organic mass spectrometry (SIOMS) has been performed on some clinically important drugs (imipramine, desipramine, clomipramine, lidocaine) by using quadrupole mass spectroscopy in which the thermal ion source has a rhenium oxide emitter. The mass spectra were presented, interpreted in a purely empirical way, by means of evidence from previous investigations, and then compared to conventional EI techniques. Sensitivity and selectivity have also been studied, demonstrating that (a) these drug compounds are efficiently surface-ionized, (b) experimental results rationalize the high sensitivity of the surface ionization detector (SID) of gas chromatography (GC) for the examined drugs, and (c) the GC/SIOMS coupling can be used for sensitive and selective detection of the drugs in the serum. An approach to detection of these drugs in serum by GC/SIOMS and GC with SID is described. The characteristics of both methods provide a reliable, sensitive, and selective method, which is needed for low concentration level measurements in complex mixtures.
通过使用热离子源带有氧化铼发射极的四极质谱法,对一些临床重要药物(丙咪嗪、去甲丙咪嗪、氯米帕明、利多卡因)进行了表面电离有机质谱分析(SIOMS)。给出了质谱图,通过先前研究的证据以纯粹经验性的方式进行了解释,然后与传统的电子电离(EI)技术进行了比较。还研究了灵敏度和选择性,结果表明:(a)这些药物化合物能有效地进行表面电离;(b)实验结果解释了气相色谱(GC)的表面电离检测器(SID)对所检测药物具有高灵敏度的原因;(c)GC/SIOMS联用可用于血清中药物的灵敏和选择性检测。描述了一种通过GC/SIOMS和带有SID的GC检测血清中这些药物的方法。这两种方法的特性提供了一种可靠、灵敏且选择性的方法,这是复杂混合物中低浓度水平测量所需要的。