Diaz-Arnold A M, Wilcox L R, Arnold M A
Department of Family Dentistry, College of Dentistry, University of Iowa, Iowa City.
J Endod. 1994 Apr;20(4):164-8. doi: 10.1016/S0099-2399(06)80327-4.
An optical system to assess blood within the pulpal cavity was developed. The system included a light-emitting diode source, human incisor teeth, and a silicon photodiode detector. An automated syringe pump was used to flow bovine blood lengthwise through the teeth. Transmitted light intensities were recorded at a collection frequency of 20 reading/min and apparent absorbance units were calculated. The effects of blood concentration, blood flow rate, and pulp chamber size were investigated. It was found that changes in blood flow rate did not cause significant changes in transmitted light intensity. Increases in blood concentration and pulp chamber size resulted in larger absorbance values. These results suggest that photoplethysmography measurements may be sensitive to the amount of blood in the pulp chamber as it reflects the pulsatile-related expansion and contraction of capillaries within the pulp tissue.
开发了一种用于评估牙髓腔内血液的光学系统。该系统包括一个发光二极管光源、人类切牙和一个硅光电二极管探测器。使用自动注射泵使牛血纵向流过牙齿。以20次读数/分钟的采集频率记录透射光强度,并计算表观吸光度单位。研究了血液浓度、血流速率和牙髓腔大小的影响。发现血流速率的变化不会导致透射光强度的显著变化。血液浓度和牙髓腔大小的增加导致吸光度值增大。这些结果表明,光电容积描记法测量可能对牙髓腔内的血液量敏感,因为它反映了牙髓组织内毛细血管与搏动相关的扩张和收缩。