Diaz-Arnold A M, Arnold M A, Wilcox L R
Department of Family Dentistry, College of Dentistry, University of Iowa, Iowa City 52242, USA.
J Endod. 1996 Jan;22(1):19-22. doi: 10.1016/S0099-2399(96)80230-5.
An in vitro, flow-through optical system was designed to measure hemoglobin (Hb) concentrations in the pulp space. The system included light-emitting diodes and a silicon photodetector positioned on opposing surfaces of human teeth. A syringe pump allowed a controlled flow of blood through the pulp chamber. The Hb concentration was computed as a nonlinear function of transmitted light intensity. Transmitted light intensities were also used as indicators of oxygenation level. Optical measurements correlated with Hb values measured by the conventional cyanmethemoglobin method (r=0.993). The mean percentage error was 5.8%, and the standard error of prediction was 0.77 g/dl for Hb concentrations ranging from 4 to 20 g/dl. Deoxygenated blood exhibited up to 31% lower transmitted intensity. Light transmission through teeth may be useful in the assessment of total Hb and blood oxygenation within the pulp chamber.