Validation of the analysis of carbamazepine and its 10,11-epoxide metabolite by high-performance liquid chromatography from plasma: comparison with gas chromatography and the enzyme-multiplied immunoassay technique.

作者信息

Martens J, Banditt P

机构信息

Department of Clinical Pharmacology, University Hospital, Magdeburg, Germany.

出版信息

J Chromatogr. 1993 Oct 22;620(1):169-73. doi: 10.1016/0378-4347(93)80067-e.

Abstract

An improved procedure for the determination of carbamazepine (CBZ) and its main active metabolite carbamazepine-10,11-epoxide (CBZ-e) in serum is described. The validation of this procedure shows that the limits of detection are lower than 80 ng/ml for CBZ and 7 ng/ml for CBZ-e. The relative standard deviation does not exceed 4.7% for both compounds in an inter-day precision test. In the intra-day precision test the relative standard deviations are 1.9% for CBZ-e and 1.1% for CBZ. A comparison of the described procedure with gas chromatography and the enzyme-multiplied immunoassay technique shows good agreement, but gas chromatography seems to have a relative poor reliability.