Accuracy and cross-sensitivity of 10 different anesthetic gas monitors.

作者信息

Walder B, Lauber R, Zbinden A M

机构信息

Institute for Anaesthesiology and Intensive Care, University Hospital, Bern, Switzerland.

出版信息

J Clin Monit. 1993 Nov;9(5):364-73. doi: 10.1007/BF01618679.


DOI:10.1007/BF01618679
PMID:8106890
Abstract

OBJECTIVE: The objective of this study was to test the accuracy and cross-sensitivity of commercially available anesthetic gas monitors. METHODS: Using gas chromatography (GC) as a reference method, the accuracy, cross-sensitivity, and ability to recognize an erroneously selected agent were determined in the following 10 monitors for volatile anesthetics: Datex Capnomac Ultima-S, Datex Capnomac, Ohmeda 5330 agent monitor, Iris Dräger, Andros Dräger PM 8020 (all monochromatic, infrared analyzers), Nellcor N-2500E, Criticare POET II, Irina Dräger (all polychromatic, infrared analyzers), Siemens Servo Gas Monitor 120 (a piezoelectric analyzer), and Brüel & Kajer Type 1304 (a photoacoustic analyzer). Accuracy was determined at 0.5, 1, 2, and 4 times the minimal alveolar concentration (MAC) of either halothane or isoflurane in oxygen (O2). The cross-sensitivity tests were performed with 70 vol% nitrous oxide in O2, 5 vol% carbon dioxide in O2, 0.032 vol% alcohol in O2, and 70% water vapor in O2. The photoacoustic analyzer showed a higher accuracy for isoflurane than the polychromatic infrared monitors. The greatest inaccuracy with isoflurane was found in the Iris Dräger monitor, which had a maximal bias percentage by volume (vol%) of 0.09 at 0.5 MAC. (This bias was within the manufacturer's specified tolerance of +/- 0.1 vol% or 10% relative difference of reading, whichever is greater.) Irina Dräger was the most accurate analyzer with halothane (mean % bias [relative %] +/- SD, 0.9 +/- 2.0%). The greatest bias with halothane was found in the monochromatic infrared analyzers, with a maximal % bias at 0.5 MAC of 50.3% of the GC reading (12.4% with a new inner Nafion tube) found in the Datex Ultima monitor. The Siemens gas monitor showed a cross-sensitivity for water vapor (-0.248 vol%). The monochromatic infrared analyzers showed a small sensitivity to alcohol (additional deviation of 0.011 to 0.147 vol% at 2 MAC isoflurane) but no sensitivity to nitrous oxide. No cross-sensitivity was found in the polychromatic infrared and photoacoustic analyzers. An incorrect selection of anesthetic agent when using a monochromatic infrared analyzer can be fatal; for example, when using halothane and selecting isoflurane the values measured by the Datex Capnomac monitor were nearly 6 times: below the actual value (i.e., 1 vol% "isoflurane" on the display = 6 vol% halothane in reality). CONCLUSIONS: The photoacoustic measurement principle is more accurate than the other methods, although the polychromatic infrared analyzers are safer because they detect erroneously selected agents.

摘要

相似文献

[1]
Accuracy and cross-sensitivity of 10 different anesthetic gas monitors.

J Clin Monit. 1993-11

[2]
Evaluation of three transportable multigas anesthetic monitors: the Bruel & Kjaer Anesthetic Gas Monitor 1304, the Datex Capnomac Ultima, and the Nellcor N-2500.

J Clin Monit. 1993-4

[3]
Laboratory evaluation of the vital signs (ICOR) piezoelectric anesthetic agent analyzer.

J Clin Monit. 1991-4

[4]
Infrared analysis of volatile anesthetics: impact of monitor agent setting, volatile mixtures, and alcohol.

J Clin Monit. 1990-7

[5]
Estimation of desflurane concentration using isoflurane channel in optical infrared analyzer.

Can J Anaesth. 1999-2

[6]
Measurement of anesthetics in blood using a conventional infrared clinical gas analyzer.

Anesth Analg. 2007-9

[7]
Carbon dioxide analysers: accuracy, alarm limits and effects of interfering gases.

Can J Anaesth. 1995-7

[8]
Methane influences infrared technique anesthetic agent monitors.

J Clin Monit Comput. 1998-2

[9]
[Monochromatic infrared halogenated gas analyzer and handling errors].

Ann Fr Anesth Reanim. 1992

[10]
[Desflurane and errors of gas selection on vaporizer analyzer].

Ann Fr Anesth Reanim. 1997

引用本文的文献

[1]
Photoacoustic gas monitoring for anesthetic gas pollution measurements and its cross-sensitivity to alcoholic disinfectants.

BMC Anesthesiol. 2019-8-9

[2]
Waste anesthetic gas exposure and strategies for solution.

J Anesth. 2018-4

[3]
Monitoring cerebral oxygenation and local field potential with a variation of isoflurane concentration in a rat model.

Biomed Opt Express. 2016-9-16

[4]
Halogenated Anesthetics Determination in Urine by SPME/GC/MS and Urine Levels Relationship Evaluation with Surgical Theatres Contamination.

J Anal Methods Chem. 2014-2-27

[5]
Analysis of oxygen, anaesthesia agent and flows in anaesthesia machine.

Indian J Anaesth. 2013-9

[6]
Metabolic discrimination of select list agents by monitoring cellular responses in a multianalyte microphysiometer.

Sensors (Basel). 2009-3-23

[7]
Neuron specific metabolic adaptations following multi-day exposures to oxygen glucose deprivation.

Biochim Biophys Acta. 2010-11

[8]
Can Modern Infrared Analyzers Replace Gas Chromatography to Measure Anesthetic Vapor Concentrations?

BMC Anesthesiol. 2008-2-8

[9]
Methane influences infrared technique anesthetic agent monitors.

J Clin Monit Comput. 1998-2

[10]
Refractive indices for volatile anesthetic gases: equipment and method for calibrating vaporizers and monitors.

J Clin Monit. 1995-5

本文引用的文献

[1]
EFFECT OF NITROUS OXIDE AND OF NARCOTIC PREMEDICATION ON THE ALVEOLAR CONCENTRATION OF HALOTHANE REQUIRED FOR ANESTHESIA.

Anesthesiology. 1964

[2]
A comparative study of halothane and halopropane anesthesia including method for determining equipotency.

Anesthesiology. 1963

[3]
A laboratory investigation of a multigas monitor for anaesthesia (EMMA).

Anaesthesia. 1982-4

[4]
Assessment of vaporizer performance in low-flow and closed-circuit anesthesia.

Anesth Analg. 1980-5

[5]
Cross-sensitivity in water vapor in the Engström EMMA.

Anesth Analg. 1984-1

[6]
Minimum alveolar anesthetic concentration: a standard of anesthetic potency.

Anesthesiology. 1965

[7]
Evaluation of the Datex 'Normac' anaesthetic agent monitor.

Anaesthesia. 1985-6

[8]
Determination of the partial pressure of halothane (or isoflurane) in blood.

Br J Anaesth. 1985-8

[9]
A laboratory investigation of two new portable gas analysers.

Int J Clin Monit Comput. 1986

[10]
Evaluation of a multigas anaesthetic monitor: the Datex Capnomac.

Anaesthesia. 1988-12

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

推荐工具

医学文档翻译智能文献检索