McPeak H, Palayiwa E, Madgwick R, Sykes M K
Nuffield Department of Anaesthetics, Radcliffe Infirmary, Oxford.
Anaesthesia. 1988 Dec;43(12):1035-41. doi: 10.1111/j.1365-2044.1988.tb05705.x.
A laboratory investigation was carried out to evaluate the performance of the Datex Capnomac multigas anaesthetic agent analyser, with particular emphasis on accuracy, response and delay times, zero and gain stability and interference from water vapour. The analysis of anaesthetic vapours was less accurate than the analysis of CO2, O2 and N2O, but acceptable for clinical use. The response to square wave changes in gas composition was accurate at frequencies up to 60 per minute for CO2 and 30 per minute for O2, but with N2O and the anaesthetic vapours there was a decrease in accuracy at frequencies above 20 breaths/minute. The instrument appeared to be unaffected by water vapour.
进行了一项实验室研究,以评估Datex Capnomac多气体麻醉剂分析仪的性能,特别强调准确性、响应时间和延迟时间、零点和增益稳定性以及水蒸气的干扰。麻醉蒸气的分析不如二氧化碳、氧气和一氧化二氮的分析准确,但对于临床使用来说是可以接受的。对于气体成分的方波变化,二氧化碳在每分钟60次、氧气在每分钟30次的频率下响应准确,但对于一氧化二氮和麻醉蒸气,在频率高于每分钟20次呼吸时准确性会下降。该仪器似乎不受水蒸气的影响。