Suppr超能文献

植入人体皮下组织的安培型葡萄糖传感器的响应与校准

Responses and calibration of amperometric glucose sensors implanted in the subcutaneous tissue of man.

作者信息

Pickup J C, Claremont D J, Shaw G W

机构信息

Division of Chemical Pathology, United Medical School, Guy's Hospital, London, UK.

出版信息

Acta Diabetol. 1993;30(3):143-8. doi: 10.1007/BF00572858.

Abstract

Glucose sensors based on immobilized glucose oxidase and hydrogen peroxide detection at a platinum base electrode were constructed and studied before, during and after implantation into the subcutaneous tissue of 11 non-diabetic subjects. A 75-g oral glucose load was given to elevate the blood glucose concentration. Seven of 14 sensors responded to the oral glucose administration with an increase in current and the output of the remainder was unchanged by the glucose load. Apparent subcutaneous glucose levels calculated from the pre-implantation calibration were a mean 58% of the plasma glucose values at baseline. A two-point in vivo calibration using paired current and glucose readings at baseline and at the maximum glucose and current after glucose ingestion showed a significantly reduced sensitivity in vivo compared with pre-implantation values (mean +/- SEM 52 +/- 21.5 vs 369 +/- 127 pA/mmol-1 per litre, P = 0.003). Recalibration of the subcutaneous glucose concentrations using the in vivo calibration sensitivity and extrapolated background current (I0) gave values similar to those in plasma. The sensitivity of five sensors recalibrated in vitro after explantation was also reduced compared with pre-implantation levels and not significantly different from the in vivo characteristics. Responding and non-responding sensors did not differ with respect to preimplantation I0, sensitivity or response time. However, provisional examination of some explanted sensors by scanning electron microscopy showed coating by cellular and other amorphous material in the non-functioning electrodes. We conclude that the sensitivity of glucose sensors of this design is markedly reduced, sometimes to zero, on implantation in the subcutaneous tissue of humans.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

构建了基于固定化葡萄糖氧化酶和在铂基电极上检测过氧化氢的葡萄糖传感器,并在11名非糖尿病受试者皮下组织植入前、植入过程中和植入后进行了研究。给予75克口服葡萄糖负荷以升高血糖浓度。14个传感器中有7个对口服葡萄糖给药有电流增加的反应,其余传感器的输出不受葡萄糖负荷影响。根据植入前校准计算的皮下葡萄糖表观水平平均为基线时血浆葡萄糖值的58%。使用基线时以及摄入葡萄糖后最大葡萄糖和电流时的配对电流和葡萄糖读数进行两点体内校准,结果显示与植入前值相比,体内灵敏度显著降低(平均值±标准误:52±21.5对369±127 pA/mmol⁻¹ 每升,P = 0.003)。使用体内校准灵敏度和外推背景电流(I0)重新校准皮下葡萄糖浓度,得到的值与血浆中的值相似。5个传感器在取出后体外重新校准的灵敏度也比植入前水平降低,且与体内特性无显著差异。有反应和无反应的传感器在植入前的I0、灵敏度或响应时间方面没有差异。然而,通过扫描电子显微镜对一些取出的传感器进行初步检查发现,无功能电极上有细胞和其他无定形物质覆盖。我们得出结论,这种设计的葡萄糖传感器在植入人体皮下组织后灵敏度显著降低,有时降至零。(摘要截短至250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验