Pickup J C, Shaw G W, Claremont D J
Division of Chemical Pathology, United Medical School of Guy's Hospital, London, UK.
Biosensors. 1989;4(2):109-19. doi: 10.1016/0265-928x(89)80026-4.
The major problems with existing amperometric enzyme electrodes for glucose sensing are oxygen sensitivity and output drift. The recently described miniature glucose sensors using immobilised ferrocene (dicyclopentadienyl iron) to mediate electron transfer from glucose oxidase to a base electrode are oxygen-independent but are often unstable. In this study, we test the hypothesis that the stability of ferrocene-based sensors can be markedly improved by better retention of the enzyme at the electrode. Sensors with graphite foil as the base electrode, dimethylferrocene as the mediator and carbodiimide/glutaraldehyde for enzyme immobilisation lost most of their activity when calibration curves were compared before and after 18 h operation in vitro at 37 degrees C (mean current decrease at 20 mmol/litre glucose was 91%). This sensor type with a covering cellulose membrane also lost activity (mean current decrease of 40% after 18 h). Electrodes in which the base sensor was platinum and the enzyme was covalently attached to agarose lost no activity when operated either at 700 mV without mediator (as a hydrogen peroxide detector) or at 160 mV with mediator. We conclude that both ferrocene-mediated and hydrogen peroxide-detecting amperometric glucose sensors with a high density of covalently-linked enzyme have an operating stability in vitro which may render them suitable for clinical use as, for example, an overnight hypoglycaemia alarm.
现有的用于葡萄糖传感的电流型酶电极的主要问题是对氧气敏感和输出漂移。最近描述的使用固定化二茂铁(二环戊二烯基铁)介导电子从葡萄糖氧化酶转移到基底电极的微型葡萄糖传感器对氧气不敏感,但往往不稳定。在本研究中,我们检验了这样一个假设,即通过更好地将酶保留在电极上,可以显著提高基于二茂铁的传感器的稳定性。以石墨箔为基底电极、二甲基二茂铁为媒介体并用碳二亚胺/戊二醛固定酶的传感器,在37℃体外操作18小时前后比较校准曲线时,失去了大部分活性(在20毫摩尔/升葡萄糖时平均电流下降91%)。这种带有覆盖纤维素膜的传感器类型也失去了活性(18小时后平均电流下降40%)。基底传感器为铂且酶共价连接到琼脂糖的电极,在无媒介体时以700毫伏操作(作为过氧化氢检测器)或有媒介体时以160毫伏操作时均未失去活性。我们得出结论,具有高密度共价连接酶的二茂铁介导型和过氧化氢检测型电流型葡萄糖传感器在体外具有操作稳定性,这可能使其适合临床应用,例如作为夜间低血糖警报器。