Gutkin V I, Osipovich L A
Med Tekh. 1993 Nov-Dec(6):37-8.
The paper provides the author's decompression linearizer of indirect blood pressure measuring device which is intended for automatic maintenance of constant decreases of air pressure in the cuff and represents a controlled pneumo-resistance whose values are automatically changed under the action of pressure and air consumption in the meter cuff. The use of the linearizer enhances the accuracy of measurements and their repeatability and significantly reduces the operating time of blood pressure measurements both under routine conditions when blood pressure is measured by the Korotkov method and particularly in the auto mode.
本文介绍了作者为间接血压测量设备设计的减压线性化器,该设备旨在自动维持袖带内气压的恒定下降,它是一种受控的气动阻力,其值会在血压计袖带中的压力和空气消耗作用下自动变化。使用线性化器可提高测量的准确性及其可重复性,并显著缩短常规条件下(采用柯氏音法测量血压时)以及特别是自动模式下的血压测量操作时间。