Abstract

During flow cytometric analysis of 29 samples of human peripheral blood mononuclear cells, we observed that a light scatter-based lymphocyte gate would fail to cover all TCR (CD3)-expressing cells. A mean of 7.9 +/- 5.3% of all CD3+ cells was found outside the gate. In addition, this shifted cell subset showed an altered CD4/CD8 ratio (8.1 +/- 5.0% of all CD4+ and 13.7 +/- 5.8% of all CD8+ cells). The use of standard light scatter-based lymphocyte gates in acquisition and/or analysis of flow cytometry (FCM) data may thus lead to erroneous results, particularly if T cells are to be studied.

摘要

在对29份人类外周血单核细胞样本进行流式细胞术分析时,我们观察到基于光散射的淋巴细胞门无法覆盖所有表达TCR(CD3)的细胞。发现平均7.9±5.3%的所有CD3+细胞位于该门之外。此外,这个偏移的细胞亚群显示出CD4/CD8比例改变(所有CD4+细胞的8.1±5.0%和所有CD8+细胞的13.7±5.8%)。因此,在流式细胞术(FCM)数据的采集和/或分析中使用基于标准光散射的淋巴细胞门可能会导致错误结果,尤其是在研究T细胞时。