An in vitro study of femoral intramedullary pressures during hip replacement using modern cement technique.

作者信息

Song Y, Goodman S B, Jaffe R A

机构信息

Division of Orthopaedic Surgery, Stanford University School of Medicine, California.

出版信息

Clin Orthop Relat Res. 1994 May(302):297-304.

PMID:8168317
Abstract

Five femora (four cadaveric and one plastic) were used to measure the intramedullary pressures simultaneously at two different locations along the proximal femur during the insertion of bone cement and the femoral component using modern cement technique. The pressures were monitored by transducers located at the midpoint of each femoral stem (P1), and just beyond the tip of the femoral stem proximal to a cement plug (P2). Transient increases in intramedullary pressure were noted during the initial compaction of the bone cement using a conventional device. However, during insertion of the femoral component, the pressures at P1 and P2 increased dramatically to peak pressures exceeding 2385 mm Hg at P1 and 3710 mm Hg at P2 respectively. These pressure elevations were not sustained; eight to 10 minutes after prosthesis insertion, the pressures decreased to below baseline levels in all five femora. This probably resulted from contraction of the cement during the curing phase. Transient elevations of intramedullary pressure to levels greater than 100 times capillary pressure are produced during hip replacement using modern cement technique. The highest pressures are generated during insertion of the femoral component rather than during the cement compaction step. These findings suggest that the use of a cement compactor to improve intrusion of the cement into bone is probably unnecessary.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验