一种用于记录动物视网膜电图电位的新系统。

A new system for recording electroretinographical potentials in animals.

作者信息

Bredehorn T, Duncker G, Kuper H, Papst N

机构信息

Department of Ophthalmology, University of Kiel, FRG.

出版信息

Ophthalmologica. 1994;208(2):98-104. doi: 10.1159/000310461.

Abstract

Electroretinography can provide meaningful evidence of oculotoxic changes. We describe a new recording system specially constructed for animal experiments. It meets the standards for electroretinographical investigations and recording systems set up by international committees. The recording apparatus has a resolution of 0.49 microV (amplitude) and 1 ms with a sensitivity range of +/- 1 mV full scale and a band-pass of 0.03 to 1,000 Hz. The signals are digitized and processed by a computer. The calibration symbol is faithfully represented as a rectangular impulse with an amplitude of 100 microV and a period of 100 ms. Normal values were calculated for 25-week-old albino Wistar rats. The normal range is defined as the mean +/- 2 SD for A and B wave parameters. These normal values can serve as reference values for future studies in oculotoxicity.

摘要

视网膜电图可以提供眼毒性变化的有意义证据。我们描述了一种专门为动物实验构建的新记录系统。它符合国际委员会制定的视网膜电图研究和记录系统标准。该记录设备的分辨率为0.49微伏(幅度)和1毫秒,灵敏度范围为满量程±1毫伏,带通为0.03至1000赫兹。信号由计算机数字化处理。校准符号如实地表示为幅度为100微伏、周期为100毫秒的矩形脉冲。计算了25周龄白化Wistar大鼠的正常值。正常范围定义为A波和B波参数的平均值±2个标准差。这些正常值可作为未来眼毒性研究的参考值。