Suppr超能文献

使用内置发光二极管接触镜电极和低成本定制仪器监测轻度缺血的效果。

Monitoring the effect of mild ischemia with a built-in light-emitting diode contact lens electrode and a low-cost custom-made apparatus.

机构信息

Department of Physiology and Pharmacology, Sackler Faculty of Medicine, Tel-Aviv University, Tel-Aviv, Israel.

出版信息

Physiol Meas. 2012 Jun;33(6):1053-2. doi: 10.1088/0967-3334/33/6/1053. Epub 2012 May 4.

Abstract

Electroretinography (ERG) is widely used in clinical work and research to assess the retinal function. We evaluated an easy to build ERG setup adapted for small animals comprising two contact lens electrodes with a built-in light-emitting diode and a custom-made amplification system. The system's sensitivity was tested by monitoring ERG in albino rat eyes subjected to mild ischemia. Flash ERG was recorded by two contact lens electrodes positioned on the rat's corneas and used alternately as test or reference. The a- and b-wave amplitudes, a-wave latency, b-wave implicit time and oscillatory potentials (OPs) were analyzed. Ischemia was achieved by elevating the intraocular pressure in the eye's anterior chamber. ERG was recorded on post-ischemia (PI) days -1, 1, 3 and 7. Morphological changes were analyzed on hematoxylin/eosin stained 5 µm sections of control 7d PI retinas. In control eyes, ERG exhibited a pattern similar to a standard recording. Retinas subjected to mild ischemia preserved ordered layered morphology, exhibiting approximately 30% loss of ganglion cells and no changes in gross morphology. By day 3 PI, ischemia caused an increase in the a-wave amplitude (from 34.9 ± 2.7 to 45.4 ± 4.3 µV), a decrease in the b-wave amplitude (from 248 ± 13 to 162 ± 8 µV), an increase in a-wave latency (from 11.1 ± 0.3 to 17.3 ± 1.4 ms) and b-wave implicit time (from 81.0 ± 1.6 to 90.0 ± 2.5 ms), and attenuation of OPs. The described setup proved sensitive and reliable for evaluating subtle changes in the retinal function in small animals.

摘要

视网膜电图(ERG)在临床工作和研究中广泛用于评估视网膜功能。我们评估了一种易于构建的 ERG 设备,该设备适用于小动物,包括两个带有内置发光二极管的接触镜电极和一个定制的放大系统。通过监测轻度缺血的白化大鼠眼睛中的 ERG 来测试系统的灵敏度。闪光 ERG 通过定位在大鼠角膜上的两个接触镜电极记录,并交替用作测试或参考。分析了 a-和 b-波幅度、a-波潜伏期、b-波隐含时间和振荡电位(OP)。通过升高眼前房的眼内压来实现缺血。在缺血后(PI)第 1、1、3 和 7 天记录 ERG。在苏木精/伊红染色的 5 µm 控制 7d PI 视网膜切片上分析形态变化。在对照眼中,ERG 表现出类似于标准记录的模式。轻度缺血的视网膜保持有序分层的形态,表现出大约 30%的神经节细胞丢失,并且大体形态没有变化。在第 3 天 PI 时,缺血导致 a-波幅度增加(从 34.9 ± 2.7 到 45.4 ± 4.3 µV),b-波幅度降低(从 248 ± 13 到 162 ± 8 µV),a-波潜伏期延长(从 11.1 ± 0.3 到 17.3 ± 1.4 ms)和 b-波隐含时间延长(从 81.0 ± 1.6 到 90.0 ± 2.5 ms),以及 OP 衰减。所描述的设置被证明对评估小动物视网膜功能的细微变化敏感且可靠。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验