Jones D C, Melchior G W, Reeves M J
J Lipid Res. 1976 May;17(3):273-7.
The quantitative analysis of individual bile acids by gas-liquid chromatography has been improved by column oven temperature programming and by a new liquid phase, SP-2401. The method is fast; bile acids are well resolved; retention times are reproducible; detector responses are linear and sensitive to 0.1 mug: and there is little adsorption onto the liquid phase. The method has been successfully used for bile, and it has the potential for use on serum.