Suppr超能文献

同步左心房 - 主动脉旁路手术期间的左心室力学

Left ventricular mechanics during synchronous left atrial-aortic bypass.

作者信息

Kawaguchi O, Sapirstein J S, Daily W B, Pae W E, Pierce W S

机构信息

Department of Surgery, Pennsylvania State University, College of Medicine, Milton S. Hershey Medical Center, Hershey 17033.

出版信息

J Thorac Cardiovasc Surg. 1994 Jun;107(6):1503-11.

PMID:8196396
Abstract

The purpose of this study was to analyze left ventricular mechanics during asynchronous, pulsatile left atrial-aortic bypass before and after microsphere injection with the pressure-volume relationship. In 14 anesthetized Holstein calves, left ventricular pressure was measured with a micromanometer and ultrasonic dimension transducers measured left ventricular orthogonal diameters. Ellipsoidal geometry was used to calculate simultaneous left ventricular volume. Contractility index, pressure-volume area, external work, and potential energy were calculated during steady-state contractions. These measurements were repeated during pulsatile left atrial-aortic bypass. To induce heart failure, we injected microspheres into the left main coronary artery, and the protocol for baseline and pulsatile left atrial-aortic bypass was repeated. Despite the significant differences in the baseline contractility index (7.4 +/- 0.7 mm Hg/ml versus 4.7 +/- 0.5 mm Hg/ml), contractility index remained the same during pulsatile left atrial-aortic bypass in control and heart failure modes, respectively. Pulsatile left atrial-aortic bypass significantly decreased end-diastolic volume (22% and 17%), pressure-volume area (58% and 48%) and external work (74% and 69%, all p < 0.05) during control and heart failure measurements, respectively. However, it did not change end-systolic volume or potential energy. In conclusion, asynchronous pulsatile left atrial-aortic bypass did not affect left ventricular contractile state in either the normal or failing heart. Although decreased pressure-volume area accounts for the reduction in myocardial oxygen consumption, unchanged potential energy suggested a limited unloading of the ventricle.

摘要

本研究的目的是通过压力-容积关系分析微球注射前后非同步搏动性左心房-主动脉旁路转流期间的左心室力学。在14只麻醉的荷斯坦犊牛中,用微测压计测量左心室压力,并用超声尺寸换能器测量左心室正交直径。采用椭圆几何法计算左心室同步容积。在稳态收缩期间计算收缩性指数、压力-容积面积、外部功和势能。在搏动性左心房-主动脉旁路转流期间重复这些测量。为诱发心力衰竭,我们将微球注入左冠状动脉主干,并重复基线和搏动性左心房-主动脉旁路转流的方案。尽管基线收缩性指数存在显著差异(分别为7.4±0.7 mmHg/ml和4.7±0.5 mmHg/ml),但在对照和心力衰竭模式下,搏动性左心房-主动脉旁路转流期间收缩性指数分别保持不变。搏动性左心房-主动脉旁路转流在对照和心力衰竭测量期间分别显著降低舒张末期容积(22%和17%)、压力-容积面积(58%和48%)和外部功(74%和69%,所有p<0.05)。然而,它并未改变收缩末期容积或势能。总之,非同步搏动性左心房-主动脉旁路转流在正常心脏或衰竭心脏中均不影响左心室收缩状态。尽管压力-容积面积降低解释了心肌氧耗的减少,但势能不变表明心室卸载有限。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验