Suppr超能文献

兔体内流量触发与压力触发按需流量系统呼吸吸气功的比较。

Comparison of inspiratory work of breathing between flow-triggered and pressure-triggered demand flow systems in rabbits.

作者信息

Nishimura M, Imanaka H, Yoshiya I, Kacmarek R M

机构信息

Department of Anesthesiology, Osaka University Medical School, Japan.

出版信息

Crit Care Med. 1994 Jun;22(6):1002-9. doi: 10.1097/00003246-199406000-00020.

Abstract

OBJECTIVES

Flow-triggered continuous positive airway pressure decreases the inspiratory work of breathing in adults when compared with pressure-triggered continuous positive airway pressure. However, the effect of flow-triggered continuous positive airway pressure on work of breathing in neonates is not known. Our objective was to determine if flow-triggering was superior to pressure triggering in the presence of narrow endotracheal tubes, such as those tubes used in neonates.

DESIGN

Prospective evaluation using within-animal comparison of flow-triggering and pressure-triggering demand flow systems.

SETTING

The animal laboratory in a university hospital.

SUBJECTS

Six spontaneously breathing white rabbits, tracheostomized and intubated with 3- and 4-mm inner diameter endotracheal tubes.

INTERVENTIONS

The animals were connected to a ventilator through a standard respiratory circuit. The ventilator was randomly operated in the following modes: flow-triggered continuous positive airway pressure, pressure-triggered continuous positive airway pressure, flow-triggered with 5 cm H2O pressure support ventilation, and pressure-triggered with 5 cm H2O pressure support ventilation.

MEASUREMENTS AND MAIN RESULTS

Esophageal pressure, airway pressure, and flow signals were monitored. Control data were obtained while the rabbits were breathing room air through the endotracheal tube. With 3-mm inner diameter endotracheal tubes, the negative deflection of esophageal pressure during flow-triggered continuous positive airway pressure was significantly less than control; however, negative deflection of esophageal pressure during pressure-triggered continuous positive airway pressure did not significantly differ from control. The application of 5 cm H2O pressure support ventilation with flow-triggering decreased negative deflection of esophageal pressure significantly compared with flow-triggered continuous positive airway pressure, pressure-triggered continuous positive airway pressure, and control. With endotracheal tube inner diameter of 4 mm, flow-triggered continuous positive airway pressure and pressure-triggered continuous positive airway pressure did not show any differences compared to control. Negative deflection of esophageal pressure differed under all conditions except control when results with the 3-mm inner diameter endotracheal tube were compared with the 4-mm inner diameter endotracheal tube.

CONCLUSIONS

Flow-triggering is superior to pressure-triggering in the presence of a 3-mm inner diameter endotracheal tube. This difference was not clear with a 4-mm inner diameter endotracheal tube. The size of the endotracheal tube may be the most important variable in evaluating the approach used to ventilate small neonates.

摘要

目的

与压力触发的持续气道正压通气相比,流量触发的持续气道正压通气可降低成人的吸气呼吸功。然而,流量触发的持续气道正压通气对新生儿呼吸功的影响尚不清楚。我们的目的是确定在使用如新生儿所用的内径较窄的气管内导管时,流量触发是否优于压力触发。

设计

采用动物体内流量触发和压力触发需求流量系统比较的前瞻性评估。

设置

大学医院的动物实验室。

对象

6只自主呼吸的白色家兔,行气管切开并插入内径3毫米和4毫米的气管内导管。

干预措施

动物通过标准呼吸回路连接到呼吸机。呼吸机随机以以下模式运行:流量触发的持续气道正压通气、压力触发的持续气道正压通气、带5厘米水柱压力支持通气的流量触发以及带5厘米水柱压力支持通气的压力触发。

测量和主要结果

监测食管压力、气道压力和流量信号。在家兔通过气管内导管呼吸室内空气时获取对照数据。对于内径3毫米的气管内导管,流量触发的持续气道正压通气期间食管压力的负向偏移显著小于对照;然而,压力触发的持续气道正压通气期间食管压力的负向偏移与对照无显著差异。与流量触发的持续气道正压通气、压力触发的持续气道正压通气及对照相比,带流量触发的5厘米水柱压力支持通气应用时食管压力的负向偏移显著降低。对于内径4毫米的气管内导管,流量触发的持续气道正压通气和压力触发的持续气道正压通气与对照相比未显示任何差异。当比较内径3毫米和内径4毫米气管内导管的结果时,除对照外,在所有条件下食管压力的负向偏移均不同。

结论

在使用内径3毫米的气管内导管时,流量触发优于压力触发。对于内径4毫米的气管内导管,这种差异不明显。气管内导管的尺寸可能是评估用于小新生儿通气方法时最重要的变量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验