Kelly I M, Gardener J E, Brett A D, Richards R, Lees W R
Department of Diagnostic Imaging, University College of London Hospitals, Middlesex Hospital, England.
Radiology. 1994 Jul;192(1):253-9. doi: 10.1148/radiology.192.1.8208948.
To develop a system for three-dimensional (3D) ultrasonography (US) that readily interfaces with conventional equipment and to apply this system to fetal imaging.
An acquisition system based on transputer technology records from the video output of standard clinical scanners was interfaced with a position-sensing device, one component of which was attached to the US probe. The method differs from other 3D US methods as any conventional probe may be used, freehand scanning is possible, and no line-of-sight limitations operate. Such scanning was performed of 25 fetuses with gestation ages of 7-30 weeks.
Acquisition typically took 3-5 seconds. Acquisition and reconstruction were successful in 17 of 25 cases and were easiest to perform in the first and second trimester. Surface rendering revealed anatomic detail that was not depicted with two-dimensional scanning.
The technique has the potential for recording complete anatomic studies for review and visualization of detail that is difficult to perceive on conventional US scans.
开发一种易于与传统设备连接的三维(3D)超声(US)系统,并将该系统应用于胎儿成像。
基于晶片机技术的采集系统从标准临床扫描仪的视频输出进行记录,并与位置传感装置相连,该装置的一个部件附着于超声探头。该方法与其他3D US方法不同,因为可以使用任何传统探头,可行徒手扫描,且不存在视线限制。对25例孕龄为7 - 30周的胎儿进行了此类扫描。
采集通常需要3 - 5秒。25例中有17例采集和重建成功,且在孕早期和中期最易操作。表面渲染显示出二维扫描未显示的解剖细节。
该技术有潜力记录完整的解剖学研究以供回顾,并可视化传统超声扫描难以察觉的细节。