Moog R, Müller N
Institute for Transfusion Medicine, University Clinics Essen, FRG.
Vox Sang. 1993;65(2):154-5. doi: 10.1111/j.1423-0410.1993.tb02134.x.
Eight blood donors underwent paired plateletpheresis using the CS 3000 Plus blood cell separator. For each donor, two samples were run: one with the TNX-6 separation and the new 30-ml low-volume collection chamber (LVCC) and one with TNX-6 separation and the A-35 collection chamber. With the LVCC, 4.8 +/- 0.5 x 10(11) thrombocytes could be harvested, whereas 4.4 +/- 1.4 x 10(11) platelets (PLT) were collected with the A-35 chamber. The separation efficiencies were 51.7 +/- 6.6% for the LVCC and 43.8 +/- 10.7% for the A-35 chamber. The leukocyte contamination of the platelet concentrates was not statistically different.
八名献血者使用CS 3000 Plus血细胞分离机进行了配对血小板单采。对于每位献血者,进行了两次样本采集:一次采用TNX-6分离法和新的30毫升低容量采集腔(LVCC),另一次采用TNX-6分离法和A-35采集腔。使用LVCC时,可采集到4.8±0.5×10¹¹个血小板,而使用A-35腔时采集到4.4±1.4×10¹¹个血小板(PLT)。LVCC的分离效率为51.7±6.6%,A-35腔的分离效率为43.8±10.7%。血小板浓缩物中的白细胞污染无统计学差异。