PLT - 30:对百特CS - 3000 Plus型血细胞分离机血小板采集功能的一项改进。

PLT-30: an improvement for platelet collection on the Baxter CS-3000 Plus.

作者信息

Nussbaumer W, Plattner R, Schoppel W, Ruth N, Schönitzer D, Kurz M, Kollar I, Jüngling G, Wagner A, Höcker P

机构信息

Department of Transfusion Medicine, University Hospital, Innsbruck, Austria.

出版信息

J Clin Apher. 1994;9(4):236-9. doi: 10.1002/jca.2920090407.

Abstract

The CS-3000 is a well-established blood cell separator. To increase the efficiency of platelet collection, Baxter developed a new separation chamber, the TNX-6, for the follow-up model of the CS-3000, the CS-3000 Plus. To reduce the occurrence of platelet aggregates and to increase platelet yield and efficiency the collection chamber A-35 has been replaced by a new, volume-reduced chamber called PLT-30. We analyzed on the basis of 201 platelet separation procedures with the CS-3000 Plus fitted with both new chambers the efficiency of platelet collection. The platelet yield per unit (mean +/- SE) was 4.55 x 10(11) +/- 0.07 x 10(11) and the efficiency was 50.79 +/- 0.76%; only 3.98% of the platelet concentrates were below 3 x 10(11). To compare our data with results from the A-35 we enrolled 33 random donors (17 in Innsbruck, 16 in Vienna) in a controlled study. Using both chambers in a randomly selected sequence within a period of 1 month on each chamber, we evaluated platelet yield, efficiency, and the contamination with white blood cells (WBCs). We found a platelet yield per unit (mean +/- SE) of 3.74 x 10(11) +/- 0.11 x 10(11) using the A-35 and of 4.4 x 10(11) +/- 0.21 x 10(11) with the new PLT-30 (P = .0001). The efficiency of the collecting process (mean +/- SE) was higher with the PLT-30 (47.76 +/- 2.11%) than with the A-35 (40.68 +/- 1.29%). The difference is statistically significant (P = .0001).(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

CS - 3000是一款成熟的血细胞分离仪。为提高血小板采集效率,百特公司为CS - 3000的后续型号CS - 3000 Plus研发了一种新型分离腔TNX - 6。为减少血小板聚集的发生并提高血小板产量和效率,采集腔A - 35已被一种新的、体积更小的腔室PLT - 30所取代。我们基于201次使用配备了两种新腔室的CS - 3000 Plus进行的血小板分离程序,分析了血小板采集效率。每单位血小板产量(均值±标准误)为4.55×10¹¹±0.07×10¹¹,效率为50.79±0.76%;只有3.98%的血小板浓缩物低于3×10¹¹。为将我们的数据与A - 35的结果进行比较,我们在一项对照研究中招募了33名随机捐赠者(因斯布鲁克17名,维也纳16名)。在1个月内,对每个腔室以随机选择的顺序使用两种腔室,我们评估了血小板产量、效率以及白细胞(WBC)污染情况。我们发现使用A - 35时每单位血小板产量(均值±标准误)为3.74×10¹¹±0.11×10¹¹,使用新型PLT - 30时为4.4×10¹¹±0.21×10¹¹(P = 0.0001)。使用PLT - 30时采集过程的效率(均值±标准误)(47.76±2.11%)高于使用A - 35时(40.68±1.29%)。差异具有统计学意义(P = 0.0001)。(摘要截取自250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验