Near infrared spectroscopy in adults. Does the Invos 3100 really measure intracerebral oxygenation?

作者信息

Harris D N, Bailey S M

机构信息

Department of Anaesthesia, Royal Postgraduate Medical School, Hammersmith Hospital, London.

出版信息

Anaesthesia. 1993 Aug;48(8):694-6. doi: 10.1111/j.1365-2044.1993.tb07183.x.

Abstract

Does the Invos near infrared spectroscopy monitor detect intracranial oxygenation, or reflect oxygenation of extracerebral tissues? The change in Invos regional cerebral oxygenation was studied during a doubling of middle cerebral artery flow with hypercapnia. There was no significant change in regional cerebral oxygenation, which suggests that the Invos does not reflect intracerebral oxygenation, possibly because of its narrow optode separation. The use of near infrared spectroscopy in adults may require a more powerful machine.