Continuous on-line optical absorbance recording of blood volume changes during hemodialysis.

作者信息

Mancini E, Santoro A, Spongano M, Paolini F, Rossi M, Zucchelli P

机构信息

Divisione di Nefrologia e Dialisi Malpighi, Policlinico S. Orsola, Malpighi, Bologna, Italy.

出版信息

Artif Organs. 1993 Aug;17(8):691-4. doi: 10.1111/j.1525-1594.1993.tb00616.x.

Abstract

Conventional techniques that measure blood volume changes during hemodialysis are invasive, hard to reproduce, and provide only intermittent evaluations. To overcome these drawbacks, we have developed an optoelectronic instrument that estimates intradialytic blood volume percentage changes by the optical absorbance of blood. This device is based on the absorption of light transmitted through blood, which is directly related to the hemoglobin concentration. A personal computer interfaced to the device provides a continuous on-line graphic display of the hemoglobin levels and the percentage changes in blood volume. The noninvasive measurement of dialysis blood volume changes by an optical method may be helpful in detecting the appearance of severe hypovolemia that can be dangerous in critically ill patients.