External quality control performance in clinical chemistry: experience in Kenya.

作者信息

Mbiti M J, Ojwang P J, Orinda D A

机构信息

Department of Clinical Chemistry, Kenyatta National Hospital, Nairobi, Kenya.

出版信息

East Afr Med J. 1993 Apr;70(4 Suppl):16-20.

PMID:8223301
Abstract

Analysis of eleven biochemical laboratory tests was done during an International External Quality Assessment Scheme (IEQAS) in which the clinical chemistry laboratory at Kenyatta National Hospital participated. Technicon SMA II continuous flow system was used in the biochemical analyses apart from glucose which was assayed manually by the glucose oxidase method. Using the standard deviation index (SDI), twenty six percent of the results were found to be outside the two standard deviation (2SD) limit. However, when variance index score (VIS) was used, 42% of the results were found to be outliers. Overall, our laboratory performed poorly compared to other laboratories in both the IEQAS and the United Kingdom External Quality Assurance Scheme (UKEQAS). This poor performance is attributed to the use of improper equipment which is not regularly maintained, lack of diagnostic reagents, lack of quality control (QC) materials and inadequate staff training in the field of quality control.

摘要

在肯尼亚国家医院临床化学实验室参与的一项国际外部质量评估计划(IEQAS)期间,进行了11项生化实验室检测。除了采用葡萄糖氧化酶法手工检测葡萄糖外,生化分析使用Technicon SMA II连续流动系统。使用标准差指数(SDI)时,发现26%的结果超出两倍标准差(2SD)限值。然而,使用变异指数评分(VIS)时,发现42%的结果为异常值。总体而言,在IEQAS和英国外部质量保证计划(UKEQAS)中,我们实验室与其他实验室相比表现较差。这种不佳表现归因于使用未定期维护的不当设备、缺乏诊断试剂、缺乏质量控制(QC)材料以及质量控制领域的工作人员培训不足。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验