Strumpf R K, Heuser R R, Eagan J T
Department of Cardiology, Arizona Heart Institute and Foundation, Phoenix 85006.
Am Heart J. 1993 Nov;126(5):1204-10. doi: 10.1016/0002-8703(93)90676-z.
Intracoronary stents were designed to improve balloon dilation results; however, to accomplish this, various criteria of optimal stent deployment must be met. Standard imaging techniques are neither sensitive nor specific enough for intraprocedural use in the verification of these implantation parameters. To assess the usefulness of angioscopy in the procedural and follow-up evaluations of Palmaz-Schatz and Strecker coronary stent deployment, 17 patients underwent angioscopy, 15 during stent placement and 2 during follow-up for stent restenosis. In the latter cases, thrombus formation was suspected; however, angioscopy showed tissue subtotally occluding the lumen without thrombus, so thrombolytic therapy was avoided. Similarly, among the 15 intraprocedural assessments, angioscopy disclosed intravascular thrombus unappreciated on angiography in two cases; another patient at high risk for intravascular thrombus was found not to have clot. In four patients, angioscopy disclosed residual narrowing in need of redilation at the Palmaz-Schatz stent articulation site. Thus in 9 (53%) of 17 stent patients, angioscopic findings either guided therapeutic selection or significantly modified the anticipated procedure. Angioscopy offers important information critical to the accurate placement and evaluation of intracoronary stents.