Korytowski W, Bachowski G J, Girotti A W
Department of Biochemistry, Medical College of Wisconsin, Milwaukee 53226.
Anal Biochem. 1993 Aug 15;213(1):111-9. doi: 10.1006/abio.1993.1393.
High-performance liquid chromatography (HPLC) with reductive mode electrochemical detection on a mercury drop has been employed for the separation and determination of lipid hydroperoxides. Under the conditions used, baseline separation is achieved for three cholesterol hydroperoxide (ChOOH) standards, not only from one another, but also from two different phosphatidylcholine hydroperoxide (PCOOH) standards. Applying this method to a test system, photodynamically treated murine leukemia cells, we have identified and quantified a major family of overlapping PCOOHs and three ChOOHs, two of which are characteristic singlet oxygen adducts. In a typical separation, the detection limit is < 0.5 pmol for ChOOHs and < 50 pmol for more slowly eluting PCOOHs. In this respect, mercury drop detection outperforms all previously described electrochemical detection methods for lipid hydroperoxides and compares favorably with other HPLC-based approaches. However, in terms of equipment cost, relative simplicity of operation, and fewer potential artifacts, this method has a clear advantage over all other existing high-sensitivity methods.
采用带有汞滴还原模式电化学检测的高效液相色谱法(HPLC)对脂质氢过氧化物进行分离和测定。在所使用的条件下,三种胆固醇氢过氧化物(ChOOH)标准品不仅彼此之间实现了基线分离,而且与两种不同的磷脂酰胆碱氢过氧化物(PCOOH)标准品也实现了基线分离。将该方法应用于一个测试系统,即光动力处理的小鼠白血病细胞,我们已鉴定并定量了一个主要的重叠PCOOH家族和三种ChOOH,其中两种是典型的单线态氧加合物。在典型的分离中,ChOOH的检测限<0.5 pmol,洗脱较慢的PCOOH的检测限<50 pmol。在这方面,汞滴检测优于所有先前描述的脂质氢过氧化物电化学检测方法,并且与其他基于HPLC的方法相比具有优势。然而,在设备成本、操作相对简便性以及潜在假象较少方面,该方法相对于所有其他现有的高灵敏度方法具有明显优势。