Rockborn P, Olsson S S
Department of Orthopaedics, Norrköping Hospital, Sweden.
J Bone Joint Surg Br. 1993 Nov;75(6):865-8. doi: 10.1302/0301-620X.75B6.8245072.
We reviewed at a minimum elapsed time of five years a consecutive series of 143 primary Exeter hip replacements in which matt-surfaced femoral stems had been used. Twenty-five patients had died and six stems and two sockets had been revised before follow-up. The remaining 110 hips were all examined clinically and radiographically. In 15 hips there were radiographic signs of definite loosening of the stem and in eight suspected loosening. The acetabulum was loose in four hips. In another eight hips localised bone resorption was present without signs of loosening. Half the patients with loosening or localised bone resorption had mild pain or no pain at all. The late complication rate with the matt-surfaced Exeter femoral stem is unacceptably high.