Optimal application of high-frequency ventilation in infants: a theoretical study.

作者信息

Ghazanshahi S D, Khoo M C

机构信息

Department of Electrical Engineering, California State University, Fullerton 92634.

出版信息

IEEE Trans Biomed Eng. 1993 Aug;40(8):788-96. doi: 10.1109/10.238463.

DOI:10.1109/10.238463
PMID:8258445
Abstract

A recent multicenter study of preterm infants concluded that high-frequency ventilation (HFV) applied at 15 Hz, in comparison with conventional mechanical ventilation (CMV), did not lead to reduced incidence of barotrauma, contrary to previous expectations. The primary goal of the present theoretical study was to determine whether computed estimates of lung pressures during HFV and CMV are consistent with these findings. An existing theoretical model of lung mechanics and gas transport in HFV was modified for applicability to neonates. New features, such as expiratory flow limitation and pulmonary air leak, were also incorporated. Simulations with the model were conducted assuming combinations of frequency and tidal volume that maintained a constant level of eucapnia. We found that peak alveolar pressures and the magnitude of alveolar pressure swings resulting from HFV at 15 Hz were in general comparable to those produced by CMV in healthy neonates and infants with bronchopulmonary dysplasia; peak alveolar pressures in the latter group tended to be higher with HFV than in CMV. Application of HFV at 15 Hz was even less advantageous than CMV when pulmonary air leak was also present in the infants with bronchopulmonary dysplasia. However, the model predicted the existence of an optimal range of frequencies between 2 and 4 Hz in which alveolar pressure swings and peak alveolar pressures could be minimized, and in some cases, reduced below the levels produced by CMV.

摘要

相似文献

1
Optimal application of high-frequency ventilation in infants: a theoretical study.
IEEE Trans Biomed Eng. 1993 Aug;40(8):788-96. doi: 10.1109/10.238463.
2
Lung-protective ventilation strategies in neonatology: what do we know--what do we need to know?新生儿科的肺保护性通气策略:我们知道什么——我们需要知道什么?
Crit Care Med. 2007 Mar;35(3):925-31. doi: 10.1097/01.CCM.0000256724.70601.3A.
3
[High-frequency ventilation in the newborn infant].[新生儿高频通气]
Pediatr Med Chir. 1985 Nov-Dec;7(6):801-7.
4
Pulmonary mechanics in preterm neonates with respiratory failure treated with high-frequency oscillatory ventilation compared with conventional mechanical ventilation.与传统机械通气相比,高频振荡通气治疗呼吸衰竭的早产儿的肺力学。
Pediatrics. 1991 Apr;87(4):487-93.
5
[High-frequency oscillatory ventilation as salvage strategy in the newborn infant. Spanish multicenter study. I].[高频振荡通气作为新生儿的挽救策略。西班牙多中心研究。I]
An Esp Pediatr. 1999 Mar;50(3):269-74.
6
Management of bronchopleural fistula in the adult respiratory distress syndrome.
New Horiz. 1993 Nov;1(4):512-21.
7
Relationship between dynamic respiratory mechanics and disease heterogeneity in sheep lavage injury.绵羊灌洗损伤中动态呼吸力学与疾病异质性的关系。
Crit Care Med. 2007 Mar;35(3):870-8. doi: 10.1097/01.CCM.0000257331.42485.94.
8
Comparison of different modes of high-frequency ventilation in surfactant-deficient rabbits.表面活性剂缺乏的家兔中不同高频通气模式的比较。
Pediatr Pulmonol. 1996 Oct;22(4):263-70. doi: 10.1002/(SICI)1099-0496(199610)22:4<263::AID-PPUL6>3.0.CO;2-N.
9
Theoretical study of inspiratory flow waveforms during mechanical ventilation on pulmonary blood flow and gas exchange.机械通气时吸气气流波形对肺血流和气体交换影响的理论研究
Comput Biomed Res. 1999 Aug;32(4):355-90. doi: 10.1006/cbmr.1999.1515.
10
Feasibility of weaning and direct extubation from open lung high-frequency ventilation in preterm infants.早产儿开放式肺高频通气撤机及直接拔管的可行性
Pediatr Crit Care Med. 2009 Jan;10(1):71-5. doi: 10.1097/PCC.0b013e3181936fbe.

引用本文的文献

1
Bringing back the old: time to reevaluate the high-frequency ventilation strategy.重拾旧法:是时候重新评估高频通气策略了。
J Perinatol. 2014 Jun;34(6):464-7. doi: 10.1038/jp.2014.39. Epub 2014 Mar 13.